beautypg.com

3 performance specifications – INFICON Composer Gas Concentration Controller User Manual

Page 21

background image

1 - 3

IP

N 07

4-

28

9L

Composer Operating Manual

1.3 Performance Specifications

The following sections provide the Composer’s performance specifications.

Transducer’s swept
volume

18 cc

Temperature range of
transducer

0 - 65 °C, controlled from 25 to 65 °C; heating only

Temperature range of
controller

0 - 50 °C

Transducer’s pressure
range

Š

70 - 1000 Torr (9.33 - 133.3 kPa) operational range without zero shift

Š

0 - 1520 Torr (0 - 202.7 kPa) recommended routine exposure range

Š

0-6200 Torr (0-1655 kPa) safety range

Gas flow range

0 - 1800 sccm

Transducer’s leak rate

<1X10

-9

std. cc/sec (He); primary and secondary containment vessels

Dimensions

Š

Transducer: 3.75" H X 5.5" W X 5.6" D (9.5 cm x 14 cm x 14.2 cm)

Š

Controller: 3.5" H X 8.0" W X 12.0" D (8.9 cm X 20.3 cm X 30.5 cm)

Power (optional supply)

2.0 amps at 24 V(dc)

Sensitivity

0.00011% (1.1 X10

-6

molar) TMIn (Trimethylindium) in

Hydrogen equivalent

Reproducibility

3X10