background image

8 - 10

IP

N 07

4-

28

9L

Composer Operating Manual

8.3.2 Toolbar

Figure 8-2 Toolbar

The Main Toolbar provides easy access to the two measurement modes and
other utilities. If the Toolbar is not visible on the screen, it can be turned on by
selecting Tools >> Toolbar. The function of each button on the Toolbar is
explained below in

Table 8-3

.

Table 8-3 Toolbar buttons

Menu Equivalent

Function

Sweep

This feature is not available in Setup Tool.
Sweep Mode is a diagnostic mode. It is used
to provide information about the shape and
amplitude of the resonance peak. See

section

8.3.6 on page 8-20

.

Amplitude Track

Amplitude Track is the standard operating
mode. In this mode the sensor will find and
then lock on to the resonant frequency of the
sample gas and track its movements. See

section 8.3.7 on page 8-21

.

Gas Concentration
Calculator

This feature is not available in Setup Tool. Gas
concentration calculator is an off-line utility. It
can calculate precursor concentration either
from supplied frequency (speed of sound)
information or from gas flow conditions and
vapor pressure data. See

section 8.3.12 on

page 8-27

.

Gas Properties
Database

This database conveniently holds physical
parameters of commonly used gases. Allows
user to add, remove or edit any gas entry. See

section 8.3.11 on page 8-26

.

Setup Process
Parameters

Brings up a dialog box for configuring MFC’s
range, gas properties and control loop
parameters. See

section 8.3.5 on page 8-14

.