INFICON Composer Gas Concentration Controller User Manual

Page 151

background image

7 - 9

IP

N 07

4-

28

9L

Composer Operating Manual

R(emote)
No data sent
No Response data unless there is a communication error.
Performs an action depending on the subcommand.
Note: Remote commands will not be performed if there is a fatal system error. The

communication error code 24 (can’t complete command) will be returned.

00 Start Amp Tracking
02 Zero frequency
03 Start control loop
04 Hold control value (Must be in control for this command to be meaningful)
05 Stop control loop
06 Stop tracking
07 Put control loop in bypass mode i.e. bubbler = 0,dilution = total flow
08 Lock keyboard parameter/control entry
09 Unlock keyboard parameter/control entry
10 Enter manual control (requires Bubbler and Dilution Flow parameter values)
Syntax: 0A 52 10 floatfloat chksum
Syntax Description: R10BubblerFlowDilutionFlow
11 Set zero frequency to factory default
12 Enter Manual Control using current Bubbler Flow and Dilution Flow vales.
99 Sets up the designer command. The codes (refer to

see page 7-8

) indicate which data will

be returned and in the order requested.