4.3.5.2.3 Instrument Zero Indicators
FACTORY ZERO . . . . . . . . . . . . . . Illuminates when the instrument is first
powered on and indicates that the
reference zero is still at factory settings for
the chosen carrier gas (controlled by the
molecular weight and gamma parameters
of the carrier gas) and temperature.
USER ZERO . . . . . . . . . . . . . . . . . Illuminates when the user has set the
reference zero (which overrides the
Factory Zero). See
Ú
below, and
for procedures.
4.3.5.2.4 Communication Indicators
SEND . . . . . . . . . . . . . . . . . . . . . . Illuminates whenever the instrument is
transmitting digital data.
RECEIVE . . . . . . . . . . . . . . . . . . . . Illuminates whenever the instrument is
receiving digital data.
4.3.5.2.5 Instrument and Process Status Indicators
WARNING . . . . . . . . . . . . . . . . . . . Illuminates when various warning
conditions are present. Specific conditions
causing this indicator will be displayed in
the numeric area if an error, or by pressing
5 if a warning. See
, for meanings of the
codes.
LOCKED . . . . . . . . . . . . . . . . . . . . Illuminates when the “parameter lock” dip
switch is activated, see
, or when the remote
communications “remote lock” command
has been given, see
Commands with their Subcommands,
Data, and Response Data., on page 7-5
.
⇑
. . . . . . . . . . . . . . . . . . . . . . . . . . Illuminates when the concentration or bulk
flow is increasing.
⇓
. . . . . . . . . . . . . . . . . . . . . . . . . . Illuminates when the concentration or bulk
flow is decreasing.