Composer Operating Manual
4.3.6.2 SENSOR Connector
A DB-15 connector used for connecting the Controller to the Transducer. There
are no user connections on this connector.
IN#1
SET reference zero
18
Sets a new value of
Reference Zero—will not
take action unless: (a) steady
condition is met (see K4
above), (b) the Transducer is
at temperature set point, and
(c) a zero value for the MIN
FLOW parameter has been
set.
IN#2 ** HOLD CONTROL OUTput 19
Holds CONTROL OUTput at
present voltage value **
IN#3 ** STOP CONTROL OUTput 20
Sets CONTROL OUTput to
zero volts**
IN#4 ** START CONTROL
OUTput
21
Initiates feedback control to
maintain target concentration
set point (or value on
CONCENTRATION INput)**
IN#5 ** BYPASS
22
Initiates full flow through
dilution line (requires SW#5
of CFGR2 be set)** See
IN#6
spare 1
23
IN#7
spare 2
24
GND
system chassis ground
13, 14,
15, 16, 17
** Inputs 2, 3, 4, and 5 are mutually exclusive in Concentration Control mode.
The last command to change state from off to on will be operative. Certain
transitions are not allowed: (a) Stop to Hold, (b) Bypass to Hold. Inputs 2, 3,
and 5 are inoperative in Bulk Flow mode.
Table 4-2 System I/O Connector Function (continued)