6.5 User Diagnostics

Various failure modes can be identified by (1) carefully watching the WARNING
LED located on the front panel of LED display units, or (2) decoding the err
value displayed on LCD units. Various failure modes can be identified. See

Table 6-7

.

There are three types of failure modes:

fatal errors . . . . . . . . . . . . . . . . . . no instrument function takes place. Some

communications are operative, except for
remote commands. The Error relay is open.

errors . . . . . . . . . . . . . . . . . . . . . . the instrument continues to function and

enters Stop or Bypass. Control cannot be
entered until the error has been corrected.
The Error relay is open.

warnings . . . . . . . . . . . . . . . . . . . . a temporary condition that causes one bad

reading will either correct itself or result in
an error being indicated, or certain gas
parameter values may lead to ambiguous
concentration calculations.

Table 6-7 User Diagnostics

Error
Designation/
Type

Meaning

Warning LED
Priority/
Pattern

Indication

Actions

Retry/Clear

Display and
Communic
ations
Error Code
(hex)

NOVRAM
checksum failure/
Fatal

The battery backed RAM
checksum test failed.
The parameters stored in
memory have been set to
default values.

Double Medium
Blink

Warning
LED
Error Relay
opens

Set NOVRAM to
defaults. Place
into error/idle
mode.*

Clear with
Reset

0x8000

TI Init fail/
Fatal

The settings sent to the
AtoD / DtoA chips that are
connected to the driver and
receiver did not match the
settings query.

Full On

Warning
LED
Error Relay
opens

Place into
error/idle mode.

Clear with
reset

0x4000

Temp. Init fail/
Fatal

The settings received from
the temperature AtoD
converter chip, do not
match what was sent.

Full On

Warning
LED
Error Relay
opens

Place into
error/idle mode.*

Clear with
Reset

0x2000

DSP Boot fail/
Fatal

The DSP checksum test
failed. If DSP truly failed to
boot the TI initialization will
also fail.

Full On

Warning
LED
Error Relay
opens

Place into
error/idle mode.*

Clear with
Reset

0x1000

Temp. Control
failure/
Fatal

The heater is off and the
temperature has been
rising for 10 seconds.

Rapid Blink
(~3.5 Hz, 50%
duty)

Warning
LED
Error Relay
opens

Place into
error/idle mode.*

Clear with
Reset

0x0800

* - error/idle mode - only limited communications operative. No measurement or control functions operative.
Heater is off. Error relay opens, all other relays open.