Page 121: Section, Section 4.3.6.5

4 - 25
IP
N 07
4-
28
9L
Composer Operating Manual
4.3.6.3 POWER Connector
A DB-9 connector used for connection to a 24 volt power supply. Refer to
for a full description of the power supply requirements
and pin connections.
4.3.6.4 Option Connector
A DB-9 connector used for:
Bulk Flow Control configuration: for providing Concentration read back
output and MFC read-back input.
Concentration Control configuration: to provide concentration output for a
Composer using dilution flow in the concentration control mode.
4.3.6.5 CFGR1
The first of 2 eight contact binary switch blocks used for activating various
features. The state of these switches is only read when power is first applied to
the Controller. See
for the configuration switch number descriptions
and
for the baud rates.
NOTE: Changing the state of a switch while the Controller is powered on does
not result in a change of the selected features. Switch status is only
read during the Controller’s power on sequence.
Table 4-3 Configuration Switch #1 Descriptions
SW#
FUNCTION
Details
1
transducer type
refer to
0 = off - series 19 and higher
1 = on - series 18 and lower
2
communications configuration
(see note below)
3
communications configuration
(see note below)
4
communications configuration
(see note below)
5
spare
6
always use factory set Reference Zero (0=off, 1=on) When on,
Reference Zero may not be
user set.
7
parameter lock
(0=off, 1=on) When on,
parameters may not be altered
from the front panel. However,
all parameters may be changed
through communications.
8
spare
NOTE: The supported communication is RS-232C.