6.6 Error Detection
Specific errors can be detected in one of the following ways:
1
Observing the blink pattern on the Warning LED, refer to
for a complete list of error or warning indicated.
2
On the optional LCD Display. When an error occurs, an error code will
replace the concentration output. If there is a warning (less serious
problem) the "Warning" annunciator will be displayed; by pressing key 5 a
warning code will be displayed. For example, if there is no gas in the
transducer the display will show "ER 80", because no measurable signal
was found. A complete listing of errors/warnings is given in
3
Via RS-232C; read the error log.
4
Via the Monitor Software pop up window.
6.6.1 LED Blink Patterns
Refer to
6.6.2 Error Codes
On the LCD and in the error log, errors are uniquely coded in hexadecimal with
each bit representing a different error. This way, multiple errors can be
recorded. Refer to
for error codes. Multiple errors are
decoded as follows:
Example 1:
There are two errors present.
The displayed error is 0x0084
It is caused by:
Ambiguous Gas Warning
0x0004
No Signal Error
0x0080
Error Code summed
0x0084
Example 2:
There are three errors present.
The displayed error is 0x001C
It is caused by:
Identical Gasses
0x0010
Bad Peak Shape
0x0008
Ambiguous Gasses
0x0004
Error code summed
0x001C