4.3.5.3.1 Function and Numeric Keys
The data entry and process control portion occupies the upper right portion of 
the front panel and consists of a numeric keypad and special function keys 
described below.
STOP . . . . . . . . . . . . . . . . . . . . . . In Concentration Control mode pressing
this key halts all process control activities, 
lights the STOP state indicator and sets all 
flows to zero. This key is active in both the 
operate and programming modes. This key 
is not operative in Bulk Flow Control 
configuration.
REF ZERO . . . . . . . . . . . . . . . . . . Pressing this key replaces the factory zero
with the current resonant frequency and 
lights the display’s "User Zero" message. 
The actuation of this key will be ignored 
unless the display’s STEADY indicator,
Ú
,
is illuminated. The reference zero is used 
as the basis of calibration and 
concentration calculation. Pressing this 
key when other than pure carrier gas is 
present in the transducer will result in a 
concentration offset and improper 
concentration values, refer to 
.
This function is inoperative when a 
non-zero MIN FLOW value is used, refer 
to 
section 4.3.5.2.7 on page 4-15
PROG . . . . . . . . . . . . . . . . . . . . . . Pressing this key switches the display
function between the operate modes and 
the programming modes. 
C , E
. . . . . . . . . . . . . . . . . . . After satisfactorily entering the parameter’s
value, pressing the E key will place the 
displayed value into memory and erase the 
previous value, moving the cursor to the 
next parameter. Pressing the C key will 
erase the newly constructed value for that 
parameter and display the previous value, 
the cursor will remain at the current 
parameter.