7 - 1

IP

N 07

4-

28

9L

Composer Operating Manual

Chapter 7

External Communications

7.1 Introduction

The Composer supports RS-232C with Checksum.

7.2 Communications Connection

The Instrument System currently supports RS-232C communication only.

An industry standard 9-pin D-sub-miniature connector is required to connect to
the Controller. The length of the cable is limited to fifty feet according to
published standards. The communications interface operates using the DTE
(Data Terminal Equipment) configuration. See

Figure 7-1

and

Table 7-1

for the

RS-232C connector’s pin assignments.

Figure 7-1 19-Pin Type “D” Female Connector

*Host

**IBM compatible computer connector

Table 7-1 RS-232C Connector’s Pin Connections

Pin#

Description

DB9*

DB25**

1

Not Used.

1

-

2 TXD

Data transmitted from controller.

2

3

3 RXD

Data received by controller.

3

2

4 DSR

Not Used.

4

-

5 GND

Signal ground.

5

7

6 DTR

Output from controller indicating ready for transmit.

6

6

7 CTS

Input to controller indicating stop transmitting.

7

4

8 RTS

Not used.

8

-

9 GND

Shield Ground.

9

-