beautypg.com

Dakota Ultrasonics DFX-8 plus MANUAL2 User Manual

Page 87

background image

DFX-8

Ultrasonic Thickness Gauge

83

5) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the highlighted value.

6) Press the OK key to set the edited Threshold value, and return to the

measurement screen. The Threshold edited will now be displayed in your

Hot Menu items.

Note: Alternatively, the Digit Edit Box can be displayed, and the value edited

using the

LEFT/RIGHT and UP/DOWN arrow keys on the left button wheel

by simply pressing the ENTER key following step 4 above.

7) Repeat steps 3-6 until all the THRESH values are correctly adjusted.


The Thresholds can also be accessed and adjusted from the tabbed menus.
However, this method is more tedious than making adjustments using the Hot
Menus
. The procedure for using the tabbed menus is outlined below:

Adjusting the Thresholds using the Tabbed Menus

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the GT1 menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

This manual is related to the following products: