Page 78

background image

Dakota Ultrasonics

74

7) Press the OK key to return to the measurement screen, or ESC to cancel

entering the DELAY.

8) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.


Adjusting the Range (B-Depth)

Once again, just to reiterate, the Range (B-Depth) + Delay (B-Start) equals the
right side of the screen
. Therefore, the Range (B-Depth) is the overall area, from
the delay, that will be viewable on the screen. The DFX-8 digitizer will round up from
the Range (B-Depth) that’s entered. Therefore, if the Range (B-Depth) is set at 1.0”,
the digitizer will round this value up to the next adjustment available. The procedures
to adjust the overall thickness range viewed Range (B-Depth) are outlined below:

Adjusting the Range (B-Depth) using the Hot Menus

1) Press the MEAS key once to activate measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left, until the WIDTH cell is highlighted. Alternatively, press the

LEFT and RIGHT arrow keys multiple times using the left button wheel.

2) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

scroll the highlighted value.