Dakota Ultrasonics DFX-8 plus MANUAL2 User Manual
Page 132
Dakota Ultrasonics
128
1) Press the MENU key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple times to tab left until
the TUNE menu is highlighted and displaying the submenu items.
Alternatively, press the
LEFT and RIGHT arrow keys multiple times
using the left button wheel.
2) Use the
UP and DOWN arrow keys multiple times using the left button
wheel to scroll through the sub menu items until PULSE is highlighted.
3) Use the
LEFT and RIGHT
arrow keys multiple times on the right button
wheel to scroll the Pulse options.
4) Once the Pulse is displayed, press the MEAS key to return to the
measurement screen.
10.12 Pulser Voltage
The DFX-8 has a 200 volt square wave pulser that can be adjusted for specific
applications and transducers. The Pulser Volt feature offers a 50 volt cut/boost
option to the user. The standard setting is 150 volts. This enables the DFX-8 to offer
greater penetration for difficult material types, or increased resolution on noisy
materials. The procedure to change the pulser voltage is outlined below:
Changing the Pulser Voltage