beautypg.com

Dakota Ultrasonics DFX-8 plus MANUAL2 User Manual

Page 17

background image

DFX-8

Ultrasonic Thickness Gauge

13

Using a Known Thickness

Note: Be sure that the probe zero procedure has been performed prior to

performing this calibration procedure.

1) Physically measure an exact sample of the material or a location directly on

the material to be measured using a set of calipers or a digital micrometer.

2) Apply a drop of couplant on the transducer and place the transducer in

steady contact with the sample or actual test material. Be sure that the

reading is stable and the repeatability indicator, in the top left corner of the

display, is fully lit and stable. Press the MENU key once to activate the

menu items tab. Press the MENU key multiple times to tab right and the

ESC key multiple times to tab left until the CAL menu is highlighted and

displaying the submenu items. Alternatively, press the

LEFT and

RIGHT arrow keys multiple times using the left button wheel.

3) Use the

UP and DOWN arrow keys on the left button wheel to scroll

through the sub menu items until MATL 1PT is highlighted.

This manual is related to the following products: