DFX-8

Ultrasonic Thickness Gauge

27

3.2 Probe – Menu


ZERO PROBE: The DFX-8 is zeroed in much the same way that a mechanical
micrometer is zeroed. If the DFX-8 is not zeroed correctly, all of the measurements
made using the DFX-8 may be in error by some fixed value. The DFX-8 is equipped
with an optional automatic or manual zero feature. Refer to the section on page 51,
for an explanation of this important procedure.
ZERO COATING: In order to account for very slight electronic differences in
transducers of the same type, frequency, and diameter, the DFX-8 has been
equipped with a “zero coating” feature. This enables the DFX-8 to obtain very
accurate readings on coatings, eliminating potential errors incurred from slight
differences in the manufacturing processes. Refer to the section on page 51, for a
detailed explanation.
TYPE: Enables the user to select the type of transducer being used from a chart of
transducer types. This provides increased linearity between transducers. Refer to
page 48 for a further explanation.

3.3 CAL – Menu



MAT:
Select the material velocity from a chart of basic material types when a known
sample thickness, or material velocity cannot be obtained. Refer to page 60 for
further info.
MATL 1PT:
Performs a single point calibration. This option allows the user to
automatically calculate the velocity by entering a known sample thickness. Refer to
page 57 for further info.
MATL 2PT: Performs a two-point calibration. This option allows the user to
automatically calculate the velocity by entering a second known sample thickness.
Refer to page 59 for further info.
VELOCITY: Function to calibrate the DFX-8 by setting the velocity to a known
material velocity. Refer to page 55 for further info.