3) Use the

UP and DOWN arrow keys multiple times using the left button

wheel to scroll through the sub menu items until COATING 2PT is

highlighted.

4) Press the ENTER key to display the Digits Edit Box.

5) Press the

UP and DOWN arrow keys multiple times using the left button

wheel to scroll the highlighted value.

6) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the digit locations.

7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.

8) Press the OK key to calculate the velocity and return to the menu screen, or

ESC to cancel the coating one point calibration.

9) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

Note: CHECK YOUR CALIBRATION! Place the transducer back on both

calibration points. The coating thickness readings should now match the known

thickness values of each sample. If the thickness is not correct, repeat the

steps above.