DFX-8

Ultrasonic Thickness Gauge

55

greater accuracy over small ranges by calculating the probe zero and velocity. The
DFX-8

provides four simple methods for setting the sound-velocity outlined below:

Known Velocity

If the material velocity is known, the user may wish to simply enter the velocity
number into the DFX-8, rather than have the DFX-8 calculate the velocity value using
a known thickness on a material sample. The steps for entering the velocity are
outlined below:

Using a Known Material Velocity

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the CAL menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Press the

UP and DOWN arrow keys multiple times using the left button

wheel to scroll through the sub menu items until VELOCITY is highlighted.

3) Press the ENTER key to display the Digits Edit Box.