beautypg.com

Dakota Ultrasonics DFX-8 plus MANUAL2 User Manual

Page 18

background image

Dakota Ultrasonics

14

4) Press the ENTER key to display the Digits Edit Box.

5) Press the

UP and DOWN arrow keys on the left button wheel to scroll

the highlighted value.

6) Press the

LEFT and RIGHT arrow keys on the left button wheel to scroll

the digit locations.

7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.

8) Press the OK key to calculate the velocity and return to the menu screen, or

ESC to cancel the one point calibration.

9) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

Note: CHECK YOUR CALIBRATION! Place the transducer back on the

calibration point. The thickness reading should now match the known

thickness. If the thickness is not correct, repeat the steps above.

2.7 Zero Coating

In order to account for very slight electronic differences in transducers of the same
type, frequency, and diameter, the DFX-8 has been equipped with a “zero coating”
feature. This enables the DFX-8 to obtain very accurate readings on coatings,
eliminating potential errors incurred from slight differences in the manufacturing
processes. The procedure is outlined below:

Performing a Coating Zero

This manual is related to the following products: