DFX-8

Ultrasonic Thickness Gauge

11

2) Press the OK key to perform the automatic probe zero, or ESC key to

cancel the zero operation.

3) The screens illustrated above will be briefly displayed followed by the main

measurement screen. The DFX-8 is ready to be calibrated.


Performing a Manual Probe Zero (On Block)

Note: When the zero probe option is set to manual, the probe zero disk

(battery cap) located on the top of the gauge, will be used as a zero standard

and the warning screen illustrated above will be displayed.

1) Press the OK or ESC keys to enter the main measurement screen and

begin the manual zero process.

2) Apply a drop of couplant on the transducer and place the transducer in

steady contact with the probe zero disk, and obtain a steady reading.

3) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the PRB menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel to scroll the tabbed menus.