beautypg.com

Dakota Ultrasonics DFX-8 plus MANUAL2 User Manual

Page 23

background image

DFX-8

Ultrasonic Thickness Gauge

19

corner of the display, is fully lit and stable. Press the MENU key once to

activate the menu items tab. Press the MENU key multiple times to tab right

and the ESC key multiple times to tab left until the CAL menu is highlighted

and displaying the submenu items. Alternatively, press the

LEFT and

RIGHT arrow keys multiple times using the left button wheel.

3) Use the

UP and DOWN arrow keys on the left button wheel to scroll

through the sub menu items until COATING 1PT is highlighted.

4) Press the ENTER key to display the Digits Edit Box.

5) Press the

UP and DOWN arrow keys on the left button wheel to scroll

the highlighted value.

6) Press the

LEFT and RIGHT arrow keys on the left button wheel to scroll

the digit locations.

7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.

8) Press the OK key to calculate the coating velocity and return to the menu

screen, or ESC to cancel the one point calibration.

9) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

Note: CHECK YOUR CALIBRATION! Place the transducer back on the

calibration point. The coating thickness reading should now match the known

This manual is related to the following products: