DFX-8
Ultrasonic Thickness Gauge
19
corner of the display, is fully lit and stable. Press the MENU key once to
activate the menu items tab. Press the MENU key multiple times to tab right
and the ESC key multiple times to tab left until the CAL menu is highlighted
and displaying the submenu items. Alternatively, press the
LEFT and
RIGHT arrow keys multiple times using the left button wheel.
3) Use the
UP and DOWN arrow keys on the left button wheel to scroll
through the sub menu items until COATING 1PT is highlighted.
4) Press the ENTER key to display the Digits Edit Box.
5) Press the
UP and DOWN arrow keys on the left button wheel to scroll
the highlighted value.
6) Press the
LEFT and RIGHT arrow keys on the left button wheel to scroll
the digit locations.
7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.
8) Press the OK key to calculate the coating velocity and return to the menu
screen, or ESC to cancel the one point calibration.
9) Finally, press the MEAS key to return to the measurement screen and begin
taking readings.
Note: CHECK YOUR CALIBRATION! Place the transducer back on the
calibration point. The coating thickness reading should now match the known