beautypg.com

Dakota Ultrasonics DFX-8 plus MANUAL2 User Manual

Page 62

background image

Dakota Ultrasonics

58

displaying the submenu items. Alternatively, press the

LEFT and

RIGHT arrow keys multiple times using the left button wheel.

3) Press the

UP and DOWN arrow keys multiple times using the left button

wheel to scroll through the sub menu items until MATL 1PT is highlighted.

4) Press the ENTER key to display the Digits Edit Box.

5) Press the

UP and DOWN arrow keys multiple times using the left button

wheel to scroll the highlighted value.

6) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the digit locations.

7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.

8) Press the OK key to calculate the velocity and return to the menu screen, or

ESC to cancel the one point calibration.

9) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

Note: CHECK YOUR CALIBRATION! Place the transducer back on the

calibration point. The thickness reading should now match the known

thickness. If the thickness is not correct, repeat the steps above.

At some point there may become a requirement for improved accuracy over a smaller
measurement range. In this case, a two point calibration would be most suited for
the job. For example, if the measurement range was .080” (2.03mm) to .250”
(6.35mm), the user would perform a one point calibration on a known thickness

This manual is related to the following products: