Programming examples -129, Gpib operation hp 438a emulation commands – Anritsu ML2430A User Manual
Page 214

Remarks:
Triggers a single reading which is added to the internal digital filter and the
updated filter power level is returned on the GPIB. After a TR1 command,
the instrument returns to TR0 standby mode.
Related
Commands:
TR0, TR2, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR2
Trigger with a settling delay
Syntax:
TR2
Remarks:
Triggers a new series of readings; enough to update the internal digital fil-
ter for a noise free reading at the current power level. The value is then re-
turned on the GPIB and returns to TR0 standby mode.
Related
Commands:
TR0, TR1, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR3
Trigger free run
Syntax:
TR3
Remarks:
Sets the ML2430A Series back into free run mode on both channels.
Related
Commands:
TR0, TR1, TR2, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
ZE
Zero sensors
Syntax:
ZE
Remarks:
Zero all connected sensors. The ML2430, when emulating the HP 438,
may take longer to Zero a sensor than the HP 438 itself. When performing
a zero the status byte should be used to identify when ZEROing is com-
plete.
ML2430A OM
6-127
GPIB OPERATION
HP 438A EMULATION COMMANDS