Syntax:
HOLD
state:
ON or OFF
Remarks:
This command holds the present graph displayed on the screen and is
available in all graph modes. In Profile and Power vs. Time modes, this
command will not work when trigger source is set to MANUAL. The held
graph can be requested over GPIB by using the OGD or OGBD com-
mands. The same graph data will be held until HOLD is switched off.
Related
Commands:
OGD, OGBD
Query:
HOLD?
Returned
String:
HOLD
IBBLP
Blanking active TTL level
BNC
Syntax:
IBBLP
polarity:
POS (positive, for high TTL level)
NEG (negative, for low TTL level)
Remarks:
Changes the expected polarity of the TTL Blanking input signal.
Query:
IBBLP?
Returned
String:
IBBLP
KEYCK
Turn key click sound on or off
SYSTEM
Syntax:
KEYCK
state:
ON or OFF
Remarks:
When ON, an audible annunciator produces a click corresponding to every
key press.
Query:
KEYCK?
Returned
String:
KEYCK
LINK
Trigger linking
TRIGGER
Syntax:
LINK
ML2430A OM
6-51
GPIB OPERATION