Gpib operation hp 438a emulation commands – Anritsu ML2430A User Manual
Page 204

6-15
HP 438A
EMULATION
COMMANDS
This section provides an alphabetical listing of the GPIB commands (mne-
monics) used to program the Model ML2430A Series Power Meter when
in HP 438A Emulation mode. The emulation mode can be set through the
front panel
SYSTEM|more|more|Rear panel|GPIB|MODE menu (see
Chapter 4, Operation) or through the GPIB command EMUL (see page
6-120).
HP Emulation commands must not have a space between the command
header and the parameter, or commas between the parameters.
The format for HP Emulation commands is:
n>... The end of the command text must be terminated with a line feed charac- The ML2430A Series in HP 438A emulation mode also supports the HP ?ID HP Identity request Syntax: ?ID Remarks: The format of the returned string is: @1 Set SRE mask Syntax: @1 val: 8-bit mask Remarks: Status Byte Structure, HP 438A: AD Set display to A – B Syntax: AD ML2430A OM 6-117 GPIB OPERATION HP 438A EMULATION COMMANDS
ter (0Ah, decimal 10) or a GPIB End of Transmission State (EOI), or both.
437B cal factor table edit and read commands.
Bit 0: Data ready
Bit 1: Cal/Zero complete
Bit 2: Entry Error
Bit 3: Measurement error
Bit 4: Over/Under limit
Bit 5: Event Status Register (HP 437B only)
Bit 6: Request Service
Bit 7: N/A