Hp 437b emulation commands gpib operation – Anritsu ML2430A User Manual
Page 195

Example:
To set the frequency of the input signal to 300 MHz:
FR300MZ
GT
Set group trigger
Syntax:
GT
mode:
0
1
2
Remarks:
0 = Ignore Group Execute Trigger (GET) command
1 = Trigger immediate response to ‘GET’ command
2 = Trigger with delayed response to ‘GET’ command
The GTn command configures what the device does when it receives the
‘GET’ command.
For example: GT1 sets the’ GET’ (Group Execute Trigger) to perform a TR1
type trigger.
Related
Commands:
TR
ID
Return identification string
Syntax:
ID
Remarks:
Returned format:
IDN?
HP 437B identity request
Syntax:
IDN?
Remarks:
Returned format:
KB
Calibration factor
Syntax:
KB
val:
1.0 to 150.0%
6-108
ML2430A OM
HP 437B EMULATION COMMANDS
GPIB OPERATION