beautypg.com

Gpib operation ml4803a emulation commands – Anritsu ML2430A User Manual

Page 176

background image

Set the offset value at the specified memory location in
dBm.

Syntax:

MOF<;>

mem:

Memory location 1 to 30.

value:

Offset value in dBm.

Remarks:

Set the offset value at memory store address to dBm

MRF

Set the reference value at the specified memory location in
dBm.

Syntax:

MRF<;>

mem:

Memory location 1 to 30.

value:

Reference value in dBm.

Remarks:

Set the reference value at memory store address to dBm.

ODT

Output the current calibration factor, offset value, and refer-
ence level.

Remarks:

These are output as three separate messages in the output buffer, as
shown below:

CAL factor: 18 ASCII characters +
OFFSET value: 18 ASCII characters +
REFERENCE level: 19 ASCII characters +

NOTE

When the ODT, OMR, and OPW data output commands
are received simultaneously, only the command which is
received last is valid.

ML2430A OM

6-89

GPIB OPERATION

ML4803A EMULATION COMMANDS

C A L
O F F S E T
R E F E R E N C E

d B
d B
d B m

.

.

.

5 CHARACTER DATA

10

15

5

1

Figure 6-3. ODT Data Output Format