beautypg.com

Gpib operation ml4803a emulation commands – Anritsu ML2430A User Manual

Page 174

background image

When selecting GPIB emulation modes, the instrument configures itself to
the preset conditions of the instrument to be emulated. For example, when
selecting HP 438A emulation, the front panel menus pass through the pre-
sets for the HP 437B (which presets sensor A to dBm) then selects HP
438A emulation (which presets sensor A to Watts).

NOTE

This command must be entered using the 488.2 format;
that is, EMUL ( = white space).

MCA

Set the cal factor value at the specified memory location in
dBm.

Syntax:

MCA<;>

mem:

Memory location 1 to 30.

value:

Cal factor value in dBm.

Remarks:

Set the cal factor value at memory store address to dBm.

MCC

Clears the cal factor value at the specified memory loca-
tion.

Syntax:

MCC

mem:

Memory location 1 to 30.

Remarks:

Clears the cal factor value at memory store to 0.0 dBm.

MCO

Clears the offset value at the specified memory location.

Syntax:

MCO

mem:

Memory location 1 to 30.

Remarks:

Clears the offset value at memory store to 0.0 dBm.

MCQ

Clears the frequency value at the specified memory loca-
tion.

Syntax:

MCQ

mem:

Memory location 1 to 30.

Remarks:

Clears the frequency value at memory store to 0.1MHz.

ML2430A OM

6-87

GPIB OPERATION

ML4803A EMULATION COMMANDS