Clears the cal factor table to one entry for 50 MHz at 100%, but does not clear the identity of the table. The cleared table is automatically saved to the sensor.
CFUID
Cal factor table identity update
SENSOR
Syntax:
CFUID ,
,
s:
A or B
table
number:
1 to number of tables supported by the sensor type
identity:
Seven characters or until a message terminator will be read as the identity.
Remarks:
Updates the seven character identity string. This only affects the copy of
the cal factor table stored in the memory of the power meter. To take affect and not be lost, the table must be saved to the sensor using the
CFUSAV
command.
Query:
CFUID? ,
Returned
String:
CFUID ,
,
CFUNITS
Cal factor display units
SENSOR
Syntax:
CFUNITS ,
s:
A or B
units:
% or PCT dB or dBm
Remarks:
This command changes the display units of the calfactors between either dB or pecentage. Note that this will also set the form the data is output over the GPIB (or RS232) when requested.
Query:
CFUNITS?
Returned
String:
CFUNITS ,
CFULD
Cal factor table binary load
SENSOR
Syntax:
CFULD ,
,,
s:
A or B
table
number:
1 to number of tables supported by the sensor type