Gpib operation hp 437b emulation commands – Anritsu ML2430A User Manual
Page 194

FA
Auto average
Syntax:
FA
Remarks:
Automatic Filter on. Allows the system to automatically select the filter
used to reduce the jitter in the display.
Related
Commands:
FM, FH
FH
average hold
Syntax:
FH
Remarks:
Hold filter sets the filter mode to Manual from Auto, but retains the auto fil-
ter setting. This function is the same as the AVGM command.
Related
Commands:
FM, FH, FA
FM
Set average value
Syntax:
FM
val:
1 to 512
Remarks:
Sets the filter length for the averaging of sensor data. For HP 437B emula-
tion, the command accepts 1 to 512 in 2-to-the-power steps. For example,
1, 2, 4, 8, 16,...256, 512.
Related
Commands:
FH, FA
FR
Frequency of the input signal
Syntax:
FR
val:
units:
GZ (GHz)
MZ (MHz)
KZ (KHz)
HZ (Hz)
EN (Hz)
Remarks:
Sets the frequency of the input signal so that the correct cal factor is used.
ML2430A OM
6-107
GPIB OPERATION
HP 437B EMULATION COMMANDS