Ml24xxa native commands gpib operation – Anritsu ML2430A User Manual
Page 105

same command as the one which changes to AUTO averaging, it will also
update the USER averaging number.
Example:
AVG A, AUTO, 64
This command will set the system to AUTO averaging and the USER av-
eraging number to 64. But, the Auto Averaging measurement system does
not use the USER averaging number.
NOTE
The AVG mnemonic can be sent to just change the
a number, but there must be a following COMMA to indi-
cate the
ample below.
The AVG mnemonic can also be sent to just change the
User Average Number
the averaging mode, but there must be a COMMA to indi-
cate the
example below.
Examples:
AVG A, AUTO,
Change Sensor A to Auto Averaging (note following comma to indicate the
AVG A, AUTO, 64
Change Sensor A to AUTO and the User Average Number to 64.
AVG A, MOV, 32
Change sensor A to MOVing average and the User Average number to
32.
AVG A, RPT,
Change sensor A to RePeaT average and keep the User Average number
as 32.
AVG A, , 128
Change sensor A User Average number to 128, but keep the previously
set averaging mode (note comma to indicate the
not being sent).
Query:
AVG?
Returned
String:
AVG ,
6-18
ML2430A OM
ML24XXA NATIVE COMMANDS
GPIB OPERATION