Gpib operation hp 438a emulation commands – Anritsu ML2430A User Manual

Page 208

background image

Remarks:

Hold filter sets the filter mode to Manual from Auto, but retains the auto fil-
ter setting. This function is the same as the AVGM command.

Related

Commands:

FM, FH, FA

FM

Set average value

Syntax:

FMEN

val:

0 to 9

Remarks:

Sets the filter length for the averaging of sensor data. For HP 438A emula-
tion, the filter length is defined as the number 2 to the power of . For
example, the command FM5EN would be 2

5

, or 32.

Related

Commands:

FH, FA

GT

Set group trigger

Syntax:

GT

mode:

0
1
2

Remarks:

0 = Ignore Group Execute Trigger (GET) command
1 = Trigger immediate response to ‘GET’ command
2 = Trigger with delayed response to ‘GET’ command
The GTn command configures what the device does when it receives the
‘GET’ command.
For example: GT1 sets the’ GET’ (Group Execute Trigger) to perform a TR1
type trigger.

Related

Commands:

TR

KB

Calibration factor

Syntax:

KB

val:

1.0 to 150.0%

terminator:

%
PCT
EN

ML2430A OM

6-121

GPIB OPERATION

HP 438A EMULATION COMMANDS