Gpib operation hp 438a emulation commands – Anritsu ML2430A User Manual
Page 208

Remarks:
Hold filter sets the filter mode to Manual from Auto, but retains the auto fil-
ter setting. This function is the same as the AVGM command.
Related
Commands:
FM, FH, FA
FM
Set average value
Syntax:
FM
val:
0 to 9
Remarks:
Sets the filter length for the averaging of sensor data. For HP 438A emula-
tion, the filter length is defined as the number 2 to the power of
example, the command FM5EN would be 2
5
, or 32.
Related
Commands:
FH, FA
GT
Set group trigger
Syntax:
GT
mode:
0
1
2
Remarks:
0 = Ignore Group Execute Trigger (GET) command
1 = Trigger immediate response to ‘GET’ command
2 = Trigger with delayed response to ‘GET’ command
The GTn command configures what the device does when it receives the
‘GET’ command.
For example: GT1 sets the’ GET’ (Group Execute Trigger) to perform a TR1
type trigger.
Related
Commands:
TR
KB
Calibration factor
Syntax:
KB
val:
1.0 to 150.0%
terminator:
%
PCT
EN
ML2430A OM
6-121
GPIB OPERATION
HP 438A EMULATION COMMANDS