This section provides an alphabetical listing of the commands (mnemon-
ics) used to program the Model ML2430A Series Power Meter when in HP
437B Emulation mode. The emulation mode can be set through the front
panel
SYSTEM|Rear Panel|GPIB|MODE menu (see Chapter 4, Opera-
tion) or through the GPIB command EMUL (page 6-105).
HP Emulation commands must not have a space between the command
header and the parameter, or commas between the parameters.
The format for HP Emulation commands is:
n>... The end of the command text must be terminated with a line feed charac- *CLS Clear GPIB status bytes Syntax: *CLS Remarks: This command performs a status structure clear command. The event *ESE Set the Event Status register enable mask Syntax: *ESE val: 8-bit mask Remarks: Event registers for the HP 437B (see Figure 6-8): 6-100 ML2430A OM HP 437B EMULATION COMMANDS GPIB OPERATION
ter (0Ah, decimal 10) or a GPIB End of Transmission State (EOI), or both.
status register and the status register are cleared except for the MAV bit.
Bit 7: Power ON
Bit 6: N/A
Bit 5: Command error
Bit 4: Execution error
Bit 3: Device Dependent error
Bit 2: N/A
Bit 1: N/A
Bit 0: N/A
See the HP 437B manual for details about the HP status registers.