Gpib operation ml24xxa native commands – Anritsu ML2430A User Manual
Page 156

Remarks:
Normally when the system is powered on the ML2430A Series returns to
the state it was in when it was powered off. This includes all the offset ta-
bles, calibration adjust values, etc.
If Secure is set to ON, non-volatile memory is disabled and all stored val-
ues are reset to the factory defaults when the system is powered on. As
long as this command is ON, the system will load the presets (see Appen-
dix A, Section A-3) every time it is turned ON.
Query:
SECURE?
Returned
String:
SECURE
SENMM
Sensor Measurement mode
SENSOR
Syntax:
SENMM ,
s:
A or B
mode:
DEFAULT (carrier wave)
MOD (modulated average)
CUSTOM (user configurable trigger setup mode)
Remarks:
Tells the sensor the type of signal it is expecting. This helps the sensor to
take the best measurements.
Query:
SENMM?
Returned
String:
SENMM ,
SENMO
Universal Sensor Operation Mode
SENSOR
Syntax:
SENMO ,
s:
A or B
value:
TRMS / FCW
Remarks:
Selects between using a universal power sensor in its normal operating
mode (TRMS) and its option 1 mode (FCW). FCW can only be selected if
this option is fitted in the power sensor.
Query:
SENMO?
Returned
String:
SENMO ,
ML2430A OM
6-69
GPIB OPERATION
ML24XXA NATIVE COMMANDS