Page 156

background image

Remarks:

Normally when the system is powered on the ML2430A Series returns to
the state it was in when it was powered off. This includes all the offset ta-
bles, calibration adjust values, etc.
If Secure is set to ON, non-volatile memory is disabled and all stored val-
ues are reset to the factory defaults when the system is powered on. As
long as this command is ON, the system will load the presets (see Appen-
dix A, Section A-3) every time it is turned ON.

Query:

SECURE?

Returned

String:

SECURE

SENMM

Sensor Measurement mode

SENSOR

Syntax:

SENMM ,

s:

A or B

mode:

DEFAULT (carrier wave)
MOD (modulated average)
CUSTOM (user configurable trigger setup mode)

Remarks:

Tells the sensor the type of signal it is expecting. This helps the sensor to
take the best measurements.

Query:

SENMM?

Returned

String:

SENMM ,

SENMO

Universal Sensor Operation Mode

SENSOR

Syntax:

SENMO ,

s:

A or B

value:

TRMS / FCW

Remarks:

Selects between using a universal power sensor in its normal operating
mode (TRMS) and its option 1 mode (FCW). FCW can only be selected if
this option is fitted in the power sensor.

Query:

SENMO?

Returned

String:

SENMO ,

ML2430A OM

6-69

GPIB OPERATION

ML24XXA NATIVE COMMANDS