Gpib operation hp 438a emulation commands – Anritsu ML2430A User Manual
Page 212

mode:
0
1
2
Remarks:
Relative mode permits any measurement result to be compared in dB or
percent to a reference value. Relative mode can be enabled using the cur-
rent power reading (RL1) or the previous reference level (RL2). Successive
measurements are displayed relative to this reference value. RL0 disables
relative mode.
RM
Range hold set
Syntax:
RM
val:
0 to 5
Remarks:
Set the range to
Ranging, where the range will change to take the best measurement auto-
matically.
Example:
To set the range to 3:
RM3EN
Related
Commands:
RH
RV
Service request mask value
Syntax:
RV
Remarks:
Read service request mask value. The returned string format is:
Converting the integer value into an 8-bit binary number, each bit corre-
sponds to the Service Request mask bits.
SM
Status Message
Syntax:
SM
Remarks:
Returns the status message in the format:
AAaaBBCCccDDddEFGHIJKLMNOP
where:
AA: measurement error code
aa: entry error code
BB: operating mode
ML2430A OM
6-125
GPIB OPERATION
HP 438A EMULATION COMMANDS