SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES

7-11

nonlinearity of a PRT with the temperature
coefficient of 0.00392/

°C is minute compared

with the slope change. Entering a slope
correction factor of 0.00385/0.00392 = 0.98214
as the multiplier in Instruction 16 results in a
calculated temperature which is well within the
accuracy specifications of the PRT.

PROGRAM

01:

Full Bridge (P6)
1:

1

Reps

2:

21

±2.5 mV 60 Hz Rejection
Range

3:

3

DIFF Channel

4:

1

Excite all reps w/Exchan 1

5: 2500

mV Excitation

6:

11

Loc [ Rs_Ro ]

7:

.001

Mult

8:

.02344 Offset

02:

BR Transform Rf[X/(1-X)] (P59)
1:

1

Reps

2:

11

Loc [ Rs_Ro ]

3:

50

Multiplier (Rf)

03:

Temperature RTD (P16)
1:

1

Reps

2:

11

R/Ro Loc [ Rs_Ro ]

3:

12

Loc [ Temp_C ]

4:

.98214 Mult

5:

0

Offset

7.12 PRESSURE TRANSDUCER - 4

WIRE FULL BRIDGE

This example describes a measurement made
with a Druck PDCR 930 depth measurement
pressure transducer. The pressure transducer
was ordered for use with 5 volt positive or negative
excitation and has a range of 5 psi or about 3.5
meters of water. The transducer is used to
measure the depth of water in a stilling well.

Instruction 6, 4 Wire Full Bridge, is used to
measure the pressure transducer. The high output
of the semiconductor strain gage necessitates the
use of the 25 mV input range. The sensor is
calibrated by connecting it to the CR10X and using
Instruction 6, an excitation voltage of 2500 mV, a
multiplier of 1 and an offset of 0, noting the
readings (

∗6 Mode) with 10 cm of water above the

sensor and with 334.6 cm of water above the
sensor. The output of Instruction 6 is 1000 V

s

/V

x

or millivolts per volt excitation. At 10 cm the
reading is 0.19963 mV/V and at 334.6 cm the
reading is 6.6485 mV/V. The multiplier to yield
output in cm is:

(334.6 - 10)/(6.6485-.19963) =

50.334 cm/mV/V

The offset is determined after the pressure
transducer is installed in the stilling well. The
sensor is installed 65 cm below the water level
at the time of installation. The depth of water at
this time is determined to be 72.6 cm relative to
the desired reference. When programmed with
the multiplier determined above and an offset of
0, a reading of 65.12 is obtained. The offset for
the actual measurements is thus determined to
be 72.6 - 65.12 = 7.48 cm.

The lead length is approximately 10 feet, so
there is no appreciable error due to lead wire
resistance.

PROGRAM

01:

Full Bridge (P6)
1:

1

Reps

2:

23

±25 mV 60 Hz Rejection
Range

3:

1

DIFF Channel

4:

1

Excite all reps w/Exchan 1

5: 2500

mV Excitation

6:

1

Loc [ HT_cm ]

7:

50.334

Mult

8:

7.48

Offset

FIGURE 7.12-1. Wiring Diagram for Full Bridge Pressure Transducer