beautypg.com

10 100 ohm prt in 3 wire half bridge – Campbell Scientific CR10X Measurement and Control System User Manual

Page 93

background image

SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES

7-9

FIGURE 7.10-1. 3 Wire Half Bridge Used to Measure 100 ohm PRT

7.10 100 OHM PRT IN 3 WIRE HALF

BRIDGE

The temperature measurement requirements in
this example are the same as in Section 7.9. In
this case, a three wire half bridge, Instruction 7,
is used to measure the resistance of the PRT.
The diagram of the PRT circuit is shown in Fig.
7.10-1.

As in the example in Section 7.9, the excitation
voltage is calculated to be the maximum
possible, yet allow the +25 mV measurement
range. The 10 kohm resistor has a tolerance of
±1%; thus, the lowest resistance to expect from
it is 9.9 kohms. We calculate the maximum
excitation voltage (V

x

) to keep the voltage drop

across the PRT less than 25 mV:

0.025V > V

x

115.54/(9900+115.54);
V

x

< 2.17 V

The excitation voltage used is 2.1 V.

The multiplier used in Instruction 7 is
determined in the same manner as in Section
7.9. In this example, the multiplier (R

f

/R

0

) is

assumed to be 100.93.

The 3 wire half bridge compensates for lead
wire resistance by assuming that the resistance
of wire A is the same as the resistance of wire
B. The maximum difference expected in wire
resistance is 2%, but is more likely to be on the
order of 1%. The resistance of R

s

calculated

with Instruction 7, is actually R

s

plus the

difference in resistance of wires A and B. The
average resistance of 22 AWG wire is 16.5
ohms per 1000 feet, which would give each 500
foot lead wire a nominal resistance of 8.3 ohms.

Two percent of 8.3 ohms is 0.17 ohms.
Assuming that the greater resistance is in wire
B, the resistance measured for the PRT (R

0

=

100 ohms) in the ice bath would be 100.17
ohms, and the resistance at 40

°C would be

115.71. The measured ratio R

s

/R

0

is 1.1551;

the actual ratio is 115.54/100 = 1.1554. The
temperature computed by Instruction 16 from
the measured ratio would be about 0.1

°C lower

than the actual temperature of the PRT. This
source of error does not exist in the example in
Section 7.9, where a 4 wire half bridge is used
to measure PRT resistance.

The advantages of the 3 wire half bridge are
that it only requires 3 lead wires going to the
sensor and takes 2 single-ended input
channels, whereas the 4 wire half bridge
requires 4 wires and 2 differential channels.

A terminal input module (Model 3WHB10K) can
be used to complete the circuit in Figure F.10-1.

PROGRAM

01:

3W Half Bridge (P7)
1:

1

Reps

2:

23

±25 mV 60 Hz Rejection
Range

3:

1

SE Channel

4:

1

Excite all reps w/Exchan 1

5: 2100

mV Excitation

6:

1

Loc [ Rs_Ro ]

7: 100.93

Mult

8:

0

Offset

02:

Temperature RTD (P16)
1:

1

Reps

2:

1

R/Ro Loc [ Rs_Ro ]

3:

2

Loc [ Temp_C ]

4:

1

Mult

5:

0

Offset