beautypg.com

Campbell Scientific CR10X Measurement and Control System User Manual

Page 60

background image

SECTION 3. INSTRUCTION SET BASICS

3-6

TABLE 3.9-1. Input/Output Instruction Memory and Execution Times

INSTRUCTION

INPUT
LOC.

PROG.
BYTES

EXECUTION TIME (ms)

MEASUREMENT RANGE

Code 0

1-4 or NA

5

Code 10

11-14

15

Code 20

21-24

25

Code 30

31-34

35

1

VOLT (SE)

R

15

3.0 + 40.2 * R

4.6 + 5.2R

2.0 + 2.8R

1.6 + 20.6 * R

2.1 + 2.8R

1.8 + 2.6R

3.0 + 124.6 * R

12.8 + 13.5R

11.6 + 12.3R

2.3 + 142.4 * R

14.5 + 15.2R

12.1 + 12.8R

2

VOLT (DIFF)

R

15

1.0 + 40.0 * R

0.6 + 9.1R

0.8 + 4.1R

2.2 + 20.5 * R

0.5 + 4.2R

-0.6 + 4.1R

2.2 + 125.4 * R

0.6 + 25.8R

0.0 + 21.0R

3.5 + 142.4 * R

0.5 + 29.1R

-0.2 + 24.3R

3

PULSE

R

15

1.3 + 0.9R

4

EX-DEL-SE

R

20

0.8 + 41.9 * R

4.7 + 5.3R

2.3 + 2.9R

1.4 + 22.5 * R

2.3 + 2.8R

2.1 + 2.7R

0.9 + 126.9 * R

13.1 + 13.6R

10.7 + 11.2R

2.1 + 143.4 * R

14.8 + 15.2R

12.3 + 12.9R

above w/o delay; w/delay: F + R(Tr + delay), where F is the fixed time and Tr is the time/rep from above.

(with delay)

0.8 + R *
(41.9+5*Delay)

1.4 + R *
(22.5+5*Delay)

0.9 + R *
(126.9+5*Delay)

2.1 + R *
(143.4+5*Delay)

5

AC HALF BR

R

18

(-5.7) + 73.1 * R

6.9 + 9.6R

4.5 + 4.8R

(-2.2) + 44.7 * R

4.5 + 4.7R

4.3 + 4.4R

(-17.5) + 206.9 * R

15.2 + 26.3R

12.8 + 21.5R

4.1 + 229.4 * R

17.0 + 29.6R

14.5 + 24.8R

6

FULL BR

R

18

(-3.9) + 98.1 * R

3.3 + 17.4R

3.3 + 7.8R

(-4.3) + 54.3 * R

3.3 + 7.5R

3.3 + 6.7R

(-16.3) + 268.9 * R

0.6 + 52.7R

0.7 + 43.0R

2.9 + 301.4 * R

1.9 + 59.2R

0.7 + 49.7R

7

3W HALF BR

R

18

2.7 + 121.9 * R

6.6 + 21.3R

4.2 + 11.4R

19.4 + 64.5 * R

4.1 + 11.6R

3.9 + 10.6R

4.8 + 335.5 * R

13.6 + 56.4R

10.1 + 46.5R

(-16.7) + 385.7 * R

15.3 + 63.1R

12.9 + 53.1R

8

EX-DEL-DIFF

R

20

0.3 + 42.5 * R

4.7 + 5.3R

2.3 + 2.9R

4.5 + 22.7 * R

2.4 + 2.9R

2.1 + 2.7R

3.9 + 127.0 * R

13.1 + 13.6R

10.7 + 11.2R

3.8 + 143.8 * R

14.8 + 15.3R

12.3 + 12.9R

above w/o delay; w/delay: F + 2R(Tr + delay), where F is the fixed time and Tr is the time/rep from above.

(with delay)

0.3 +
2*R*(42.5+5*Delay)

4.5+2*R*(22.7+
5*Delay)

3.9+2*R*(127.0+5*
Delay)

3.8+2*R*(143.8+5
*Delay)

9

FULL BR-MEX

R

19

(-6.0) + 327.0 * R

1.4 + 37.0R

1.5 + 17.5R

(-7.0) + 85.0 * R

1.3 + 17.1R

1.6 + 15.3R

(-8.7) + 475.0 * R

-2.3 + 107.4R

-2.2 + 87.9R

3.3 + 541.0 * R

3.4 + 118.9R

-2.0 + 101.1R

V1 on range 5, V2 on:

1.6 + 27.2R

1.5 + 17.5R

1.6 + 17.3R

1.5 + 16.4R

0.3 + 62.4R

-0.9 + 52.6R

-0.3 + 69.1R

0.2 + 59.3R

V1 on range 15, V2 on:

1.5 + 26.2R

1.6 + 16.4R

1.6 + 16.2R

1.6 + 15.3R

0.3 + 61.3R

-0.9 + 51.6R

0.3 + 68.1R

0.3 + 58.2R

10

BATT. VOLT

1

4

8.3

11

TEMP-(107)

R

15

2.3 + 7.6R

14 + 48.4R

14.7 + 53.4R

12

RH (207)

R

17

2.3 + 9.6R

13

TEMP-TC SE

R

18

11.6 + 68.3 * R

8.5 + 9.2R

5.6 + 6.8R

11.9 + 49.2 * R

6.0 + 6.8R

5.2 + 6.7R

(-2.8) + 138.5 * R

16.4 + 17.7R

13.4 + 15.2R

13.8 + 149.7 * R

18.0 + 19.4R

15.1 + 16.9R

14

TEMP-TC DIF

R

18

11.5 + 69.3 * R

4.5 + 13.5R

4.2 + 8.3R

13.0 + 49.8 * R

4.3 + 8.6R

4.1 + 8.0R

215.8 + 35.0 * R

4.0 + 30.9R

4.6 + 25.1R

13.3 + 151.1 * R

4.2 + 34.2R

4.1 + 28.8R

15

SERIAL I/O

R*#data

29

time in parameter 9

16

TEMP-RTD

R

15

1.5 + 2.0R

17

TEMP-INTERNL

1

4

6.2

18

TIME

1 or 5

7

1.9 (options 0, 1, 2) 4.2 (option 3)

19

SIGNATURE

1

4

3600

20

PORT SET

0

6

11.3

21

PULSE PORT

0

5

1.2 + 10H; where H is hundredths of a second

22

EXCIT-DEL

0

9

0.5 + delay (ms)

23

BURST

X

35

1.333R

∗ # scans; where X = smpl before trigger + # chan ∗ scans

24

CALIBRATION

19

4

2614, 25.0 when only saving results of automatic calibration.

25

READ PORTS

1

6

0.5

26

TIMER

1 or 0

4

0.4 to reset, 0.9 to load into location

27

PER.AVG.

R

19

period of signal

∗ (No. cycles + 1.5), or time limit + 2

28

VIB.WIRE

R

21

170 + period of signal

∗ (No. cycles + 1.5), or time limit + 172

29

INW PS9105

2

88

153

101

SDM-INT8

1-8

27

2.3 + 1.65

∗ No.Values+averaging interval if used

102

SDM-SW8A

R

16

4 + F

∗ R; F=function time = 1 for state, 2 for counts or cycle, 0 for signature

103

SDM-AO4

0

6

1.1 + 0.8R

104

SDM-CD16

0

6

0.3 + 1.8R

105

SDI-12 REC

X

18

sensor dependent; where X depends on SDI-12 sensor

106

SDI-12 SEN

0

10

ttt in parameter 2, this is sensor dependent

107

SDM-CSAT3

5R

7

2.5 + 2.4R

108

SDM-UDG01

1

15

155.4

109

SDMX50

0

4

1.7 + 31.1C; C = channel number

110

SDM TRIG

0

2

1.1

114

SET TIME

3-5

5

994.6

115

SET SDM BAUD

0

0

25 + 10

∗ par 1

117

READ ID

0

4

0.8