Setting for trigger point – KYORITSU 6310 Instruction Manual User Manual
Page 85
![background image](/manuals/369632/85/background.png)
4.2.2 Measurement Setting (QUALITY-Swell, Dip, INT Measurement) KEW6310
KEW6310
4.44
Setting for trigger point
Trigger to start and stop recording, when a preset threshold is exceeded, is decided based on the number of
recorded data.
Past: 0 ~ 200 (can be set by 1)
Next : 200 ~ 0 (can be set by 1)
* Default value (or after system reset) : 100
Example of Trigger Pint Setting:
Setting item
e.g.
Reference voltage
100V
Swell 110%
Hysteresis 1%
Trigger point
Past: 100, Next: 100
Start of swell
End of swell
100%
(reference
voltage
110%
(Swell)
109%
(110% - 1%)
total 201 data pts
total 201 data pts
100
data pts
100
data pts
Trigger
Point
100
data pts
100
data pts
Trigger
Point
Duration
See also other documents in the category KYORITSU Tools:
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)