KEW6310 6.4.2 Instantaneous value measurement -Over-range / Bar indication
KEW6310
6.22
Bar Indication
The calculations and measurements performed by this instrument are based on the voltage and frequency
of V1. If the value of V1 is less than 5% of the chosen range or if the frequency is not within 40 ~ 70Hz, all
the parameters (except for voltage and current ) cannot be computed and thus displayed. In such a case,
the numerical digits will be replaced by a bar indication (“- - - -”) as shown:
Zero Indication
Zero “0” is displayed when measured items exceeds following conditions.
Voltage
: Voltage Range x VT ratio x 5%
e.g. Voltage Range : 300V, VT ratio : 1 => 15V
Current
: Current Range x CT ratio x 1%
e.g. Current Range : 200A, CT ratio : 2 => 4A
“-------“ is displayed.
0 is displayed.
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)