KEW6310 16.4 Demand measurement
KEW6310
16.6
16.4 Demand measurement ( Range)
(1) Target value (DEM Target)
Display range
Fixed set value (1.000mW ~ 999.9TW)
(2) Predicted value (DEM Guess)
Display range
Same decimal point place and unit to target value
Formula
∑
×
=
interval
demand
of
beginning
from
Period
interval
Demand
DEM
DEM
GUESS
_
_
_
_
_
_
(3) Demand value (present value) (ΣDEM)
Display range
Same decimal point place and unit to target value
Formula
(
)
interval
hour
WP
DEM
1
×
+
=
Σ
where:
∑
Σ
=
Σ
DEMi
DEM
*
1
=
i
(1P2W×1)
*
2
=
i
(1P2W×2,1P3W,3P3W)
*
3
=
i
(1P2W×3,3P3W3A,3P4W)
*
4
=
i
(1P2W×4)
(4) Load factor
Display range
0.00 ~ 9999.99%
Formula
et
T
DEM
DEM
arg
Σ
16.5 Waveform measurement ( Range)
Displayed data
2 waveforms (256 points)
Scale change
0.1/ 0.2/ 0.5/ 1.0/ 2.0/ 3.0 times of rating
16.6 Harmonic measurement ( Range)
Meas. Method
PLL synchro system
Measuring range
45 ~ 65Hz
Analysis order
1 ~ 63rd
Window width
2 cycles
Window type
Rectangular
Analysis data
512 points
Analyzing rate
approx once / 2 sec
Display item
(1) Voltage, current, THD, frequency
(2) Voltage/ Rate of content/ Phase angle at each order
Save item
(1) Voltage, current, THD
(2) Voltage/ Phase angle at each order