4 ranges and over-range indications, 1 ranges, 4 ranges and over-range indications 6.4.1 ranges – KYORITSU 6310 Instruction Manual User Manual
Page 157
![background image](/manuals/369632/157/background.png)
KEW6310 6.4.1 Instantaneous value measurement -Ranges
KEW6310
6.18
6.4 Ranges and Over-range indications
6.4.1 Ranges
Ranges and decimal points for the measuring items will be automatically adjusted depending on the
settings for Voltage, Current Ranges and VT / CT ratio.
Voltage Range : V, Max digit : 4-digit
(V Range) x (VT ratio) x (120%)
Decimal point & Unit
1.8 ~ 9.999 V
9.999 V
10 ~ 99.99 V
99.99 V
100 ~ 999.9 V
999.9 V
1 ~ 9.999kV 9.999kV
10 ~ 99.99kV 99.99kV
100 ~ 9.999kV 999.9kV
1 ~ 9.999 MV
9.999 MV
10 ~ 12.0 MV
12.00 MV
Current Range : A, Max digit : 4-digit
(A Range) x (CT ratio) x (120%)
Decimal point & Unit
1.2 ~ 9.999 mA
9.999 mA
10 ~ 99.99 mA
99.99 mA
100 ~ 999.9 mA
999.9 mA
1 ~ 9.999 A
9.999 A
10 ~ 99.99 A
99.99 A
100 ~ 999.9 A
999.9 A
1 ~ 9.999kA
9.999kA
10 ~ 99.99kA
99.99kA
100 ~ 999.9kA
999.9kA
1 ~ 9.999 MA
9.999 MA
10 ~ 36.00 MA
36.00 MA