5 capacitance calculation, 8 other specifications, Kew6310 – KYORITSU 6310 Instruction Manual User Manual
Page 291: Cos θ, 1) ac power supply, 2) dc power supply, 3) battery charge
![background image](/manuals/369632/291/background.png)
KEW6310 16.7.5 Capacitance calculation
KEW6310
16.8
16.7.5 Capacitance calculation
Display item
Same to W Range (except for the change from PA to C)
Save item
(Measurement data at W Range) + (calculated capacitance value)
Formula
[
]
[ ]
F
V
f
P
k
P
C
μ
θ
θ
θ
θ
)
1
cos
1
1
cos
1
(
2
10
var
)
1
cos
1
1
cos
1
(
0
2
1
2
2
9
0
2
1
2
−
−
⎟
⎟
⎠
⎞
⎜
⎜
⎝
⎛
−
Ч
Ч
Ч
=
−
−
⎟
⎟
⎠
⎞
⎜
⎜
⎝
⎛
−
Ч
=
−
−
−
−
−
π
C
: Capacitance needs for improvement
P
: Load power [kW]
f
:
Frequency
V :
Voltage
1
cos
θ
: Measured power factor
0
cos
θ
: New power factor (target)
16.8 Other specifications
(1) AC power supply
Voltage range
AC100 ~ 240V±10%
Frequency 45
~
65Hz
Power consumption
20VA max
(2) DC power supply
Dry
battery
Rechargeable
battery
Type Alkaline
(LR6)
Ni-MH(HR-15-51)
Rated voltage
DC9V (=1.5Vx6)
DC7.2V (=1.2Vx6)
Current consumption
500mA typ.(@9V)
560mA typ.(@7.2V)
Possible measurement time
Backlight ON: 1 hour
Backlight OFF: 2 hours
(ref. at 23ºC)
Backlight ON: 2 hours
Backlight OFF: 5 hours
(ref. at 23ºC after full-charge)
(3) Battery charge
Charging voltage
approx 9V
Charging current
approx 400mA
Charging pattern
Charging pattern is as follows to control whole current consumption.
Pattern Charging
Pause
Total
charging time
I. Power ON,LCD_ON 0.7 4.3
48
II. Power ON,LCD_OFF
2.1 2.9 14
III. Power OFF 4.2
0.8
7
[min] [hour]
Start charging
Following should be completely met.
- Supply of power from AC power supply
- Selector switch is set to “Rechargeable battery” position.
- Operation to start battery charge
Finish charging
Battery charge stops if any of following is met.
(1) power from AC power supply is stopped,
(2) selector switch is set to “Dry battery” position,
(3) 48hours later from the start of battery charge,
(4) Battery voltage becomes lower than that checked at previous pause period,
(5) charging voltage is 9.5V or more (batteries are removed),
(6) specific charging cycle is exceeded.
Battery charge stops if Any of (1), (4), (5), (6) is met.