11.3.2 Power quality – Measuring procedure (Transient measurement) KEW6310
11.13
KEW6310
Timing of data recording
<
Recording at event occurrence>
<
Recording at every interval>
* Function available with ver2.00 or later.
Avg,max and min* values within
each interval are recorded.
※
Inst value : max value of 10,000 data obtained by sampling at 100us
1sec before the preset interval comes
Avg value : Avg of inst values obtained in the preset inst interval
Max value : Max inst values obtained in the preset inst interval
Min value : Min inst values obtained in the preset inst interval
Interval Interval
201 data pts
recorded
Past 100
Data pts
Next 100
Data pts
Trigger
Threshold
110%
start
Peak(max)
100%
reference Voltage
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)