6

OPERATION

88

OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

C1.7.1

limitation

set lower and upper limit for flow

speed on all outputs

min-max: -100 - +100 m/s

C1.7.2

flow direction

select flow direction

normal, reverse

C1.7.3

time constant

within set time, measurements

are averaged, displayed and sent

to current output

min-max: 000.0 - 100.0 s

C1.7.4

low flow cutoff

beneath set flow speed, zero

appears in display

min-max: 0.000 - 10.00 m/s / 0.000 - 32.81 ft/s

C1.8.2

vel. of sound

simulation of the VoS

min-max: 0.0000 - 2500.0 m/s / 0.0000 - 8202.1

ft/s

C1.9.1

error limit

with set limits, every erroneous

measurement is counted as

percentage of the measured

values

min-max: 000 - 100 %

C1.9.2

counter decrease

amount with which the counter

decreases

min-max: 00 - 99

C1.9.3

counter limit

totalized correct measurements

equal to set counter decrease,

decrease error limit by 1

min-max: 000 - 999

C1.11

diagnosis value

diagnosis of the flow

measurement

signal quality, Reynolds number

C5.1

C5.1

C5.1

C5.1

I/O Hardware

I/O Hardware

I/O Hardware

I/O Hardware

C5.1.1

terminals A

assign terminal A

for basic I/O: current output, off

for Modular I/O: free selectable 1st IO Module

C5.1.2

terminals B

assign terminal B

for basic I/O: status output, limit switch, control

input

for Modular I/O: free selectable 2nd IO Module

C5.1.3

terminals C

assign terminal C

for basic I/O: status output, limit switch, off

for Modular I/O: fixed current out

C5.1.4

terminals D

assign terminal D

for basic I/O: status output, limit switch, pulse

output, frequency output, off

for Modular I/O: fixed current out

C5.2

C5.2

C5.2

C5.2

current out A

current out A

current out A

current out A

C5.2.1

range 0-100%

set current range

min-max: 04.0 - 20.0 mA

C5.2.2

extended range

set upper current range to

extended value

min-max: 03.5 - 21.5 mA

C5.2.3

error current

after error this selected current

is set

min-max: 03.0 - 22.0 mA

C5.2.4

error condition

error in device, out of specification, application

error

C5.2.5

measurement

measurement value to current

output

(depends on pipe configuration: 1 or 2)

volume flow, mass flow, VoS, flow speed, gain,

SNR, diagnosis value, volume flow 1 or 2, VoS 1

or 2

C5.2.6

range

set measurement range from 0

to 100 %

(min-max depends on parameter settings)

C5.2.7

polarity

set polarity of current output

positive -, negative -, both polarity(ies), absolute

value

C5.2.8

limitation

set lower and upper limit for

current output

min-max: -150 - +150 %

C5.2.9

low flow cutoff

beneath set value, current output

is set to zero

min-max: 00.0 - 20.0

C5.2.10

time constant

within set time, measurements

are averaged, displayed and sent

to current output

min-max: 000.1 - 100.0

C5.2.11

special functions

for ranging

automatic range, external range, off