beautypg.com

Start-up – KROHNE OPTISONIC 6300 EN User Manual

Page 72

background image

5

START-UP

72

OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

• X9.7: Confirm or adjust the reading with the actual distance on the rail.
• X9.8: Optimization loop. Repeat steps X9.8.1...X9.8.5 until the advised mounting distance does

not change more than 0.5%.

• X9.8.1: Optimise distance?
• X9.8.2: Read the velocity of sound of the fluid
• X9.8.3: Continue?
• X9.8.4: Confirm or adjust the velocity of sound
• X9.8.5: Read the advised mounting distance and reposition the transducer
• X9.9: Read the preliminary volume flow
• X9.10: Path ready? Enter "Yes". If you have:

1 path or pipe: you are finished, proceed with X9.12
2 paths: go to X9 for the 2

nd

path

2 pipes: go to X10 for the 2

nd

pipe

• X9.12: End Installation? If you enter "No" the installation is not saved, go to X9. If you enter

"Yes" the installation is saved and the measurement screen will appear.

• Mount the cover (see section "mounting the cover" in chapter "General mechanical

installation")

.book Page 72 Tuesday, August 18, 2009 4:51 PM