Operation – KROHNE OPTISONIC 6300 EN User Manual
Page 87
OPERATION
6
87
OPTISONIC 6300
www.krohne.com
07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en
A4.2
unit
unit for HART current output
L/s, L/min,L/h, m3/s, m3/min, m3/h, m3/d,
ft3/s, ft3/min, ft3/h, gal/s, gal/min,gal/h, gal/d,
IG/s, IG/min, IG/h, IG/d, bbl/h, bbl/d, free unit
A4.3
range
range for main HART current
output
min-max: 0.00 - xxxx
(depends on
configuration)
A4.4
low flow cutoff
low flow cutoff for main HART
current output
min-max: 00.0 - 20.0
A4.5
time constant
time constant for main HART
current output
min-max: 000.1 - 100.0
A5
A5
A5
A5
digital outputs
digital outputs
digital outputs
digital outputs
only active if HART device
pulse output A, B or D, counter 1
A5.1
measurement
value for pulse output
(depends on pipe configuration: 1 or 2 pipes)
volume flow, mass flow, volume flow 1 or 2
A5.2
pulse value unit
unit for main pulse output
(unit class depends on selected measurement)
A5.3
value p. pulse
value of pulse per volume or
mass for pulse output
(min-max depends on selected measurement)
A5.4
low flow cutoff
low flow cutoff for pulse output
(min-max depends on selected measurement)
B Test
B
B
B
B
test
test
test
test
B1.7
terminal A
(depends on IO setting hardware) current output A, frequency output A, pulse
output A, status output A, limit switch A, control
input A
B1.8
terminal B
(depends on IO setting hardware) current output B, frequency output B, pulse
output B, status output B, limit switch B, control
input B
B1.9
terminal C
(depends on IO setting hardware) current output C, status output C, limit switch C
B1.10
terminal D
(depends on IO setting hardware) frequency output D, pulse output D, status
output D, limit switch D
B3.1
C number
identification of electronics
also see converter sticker;
first line: circuit board,
second line: software,
third line:calibration or production date
B3.2.1
Sensor CPU
identification of hardware and
software for flow processing
also see sensor board of electronics
B3.2.2
Sensor DSP
identification of hardware and
software for signal processing
also see sensor board of electronics
B3.2.3
Sensor driver
identification of hardware and
software for driver part
also see sensor board of electronics
B3.3
Device
identification of circuit board
serial number of circuit board, main software
version number, production date
C Setup
C
C
C
C
setup
setup
setup
setup
C1.5
extra
measurements
extra available parameters for
display or input/output
(only for dual pipe configuration: 1, 2 pipes)
on pipe 1: mass flow, flow speed, gain, SNR
on pipe 2: mass flow, flow speed, gain, SNR
C1.6.1
zero calibration
transit time offset at zero flow
cancel, default, automatic min-max: -10000 -
+10000 ps
C1.6.2
meter factor
set factor for correction of
volume flow, mass flow, flow
speed and Reynolds number
min-max: 0.500 - 2.000
C1.6.3
Reynolds
correction
set Reynolds correction for flow
profile disturbances, effective on
volume flow, mass flow
on,off
.book Page 87 Tuesday, August 18, 2009 4:51 PM