Technical data – KROHNE OPTISONIC 6300 EN User Manual

Page 113

background image

TECHNICAL DATA

8

113

OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

Control input

Control input

Control input

Control input

Function

Hold value of the outputs (e.g. for cleaning work), set value of the

outputs to "zero", counter and error reset, range change.

Start of dosing when dosing function is activated.

Operating data

Basic I/Os

Basic I/Os

Basic I/Os

Basic I/Os

Modular I/Os

Modular I/Os

Modular I/Os

Modular I/Os

Ex-i

Ex-i

Ex-i

Ex-i

Active

-

U

int

= 24 VDC

Terminals open:

U

0, nom

= 22 V

Terminals bridged:

I

nom

= 4 mA

On:

U

0

≥ 12 V with

I

nom

= 1.9 mA

Off:

U

0

≤ 10 V with

I

nom

= 1.9 mA

-

Passive

8 V ≤ U

ext

≤ 32 VDC

I

max

= 6.5 mA

at U

ext

≤ 24 VDC

I

max

= 8.2 mA

at U

ext

≤ 32 VDC

Contact closed (On):

U

0

≥ 8 V

with I

nom

= 2.8 mA

Contact open (Off):

U

0

≤ 2.5 V

with I

nom

= 0.4 mA

3 V ≤ U

ext

≤ 32 VDC

I

max

= 9.5 mA at

U

ext

≤ 24 V

I

max

= 9.5 mA at

U

ext

≤ 32 V

Contact closed (On):

U

0

≥ 3 V

with I

nom

= 1.9 mA

Contact open (Off):

U

0

≤ 2.5 V

with I

nom

= 1.9 mA

U

ext

≤ 32 VDC

I ≤ 6 mA at U

ext

= 24 V

I ≤ 6.6 mA at

U

ext

= 32 V

On:

U

0

≥ 5.5 V or I ≥ 4 mA

Off:

U

0

≤ 3.5 V or

I ≤ 0.5 mA

U

I

= 30 V

I

I

= 100 mA

P

I

= 1 W

C

I

= 10 nF

L

I

= 0 mH

NAMUR

-

Active to EN 60947-5-

6

Contact open:

U

0, nom

= 8.7 V

Contact closed (On):

I

nom

= 7.8 mA

Contact open (off):

U

0, nom

= 6.3 V with

I

nom

= 1.9 mA

Identification for

open terminals:

U

0

≥ 8.1 V with

I ≤ 0.1 mA

Identification for

short circuited

terminals:

U

0

≤ 1.2 V with

I ≥ 6.7 mA

-

Low-flow cutoff

Low-flow cutoff

Low-flow cutoff

Low-flow cutoff

On

0...±9.999 m/s; 0...20.0%, settable in 0.1% steps, separately for each

current and pulse output

Off

0...±9.999 m/s; 0...19.0%, settable in 0.1% steps, separately for each

current and pulse output

Time constant

Time constant

Time constant

Time constant

Function

Can be set together for all flow indicators and outputs, or separately

for: current, pulse and frequency output, and for limit switches and

the 3 internal counters

Time setting

0…100 seconds, settable in 0.1 second steps

.book Page 113 Tuesday, August 18, 2009 4:51 PM