beautypg.com

Operation, 3 function description – KROHNE OPTISONIC 6300 EN User Manual

Page 86

background image

6

OPERATION

86

OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

6.3 Function description

Menu No.

Display

Function description

Selection list

X

XX

X

Installation

Installation

Installation

Installation

X3

X3

X3

X3

units

units

units

units

X3.1

size

unit for dimension

mm, inch

X3.2

volume flow

unit for volume flow

L/s, L/min,L/h, m3/s, m3/min, m3/h, m3/d,

ft3/s, ft3/min, ft3/h, gal/s, gal/min,gal/h, gal/d,

IG/s, IG/min, IG/h, IG/d, bbl/h, bbl/d, free unit

X3.3

velocity

unit for flow speed and velocity of

sound (VoS)

m/s, ft/s

X3.4

density

unit for density

kg/L, kg/m3, lb/ft3, lb/gal, free unit

X3.5

viscosity

unit for viscosity

cSt, mm2/s

X5

X5

X5

X5

number of paths

number of paths

number of paths

number of paths

in case "2 paths" is selected, the

measurement results will be

averaged

1 path, 2 paths

X6.3

diameter

size for outer diameter of pipe

min-max: 20 - 4300 mm / 0.787 - 169.3 inch

X6.4

pipe material

carbon steel, stainless steel, cast iron,

aluminum, concrete, GRF/RFP, asbestos

cement, PP/PVC, acrylics, polyamide, other

X6.5

VoS pipe material

min-max: 1000.0 - 4500.0 m/s / 3280.8 - 14764

ft/s

X6.6

wall thickness

min-max: 1.000 - 200.0 mm / 0.039 - 7.874 inch

X6.7

liner material

cement, epoxy, PP, LDPE, HDPE, PTFE, rubber,

other, none

X6.8

VoS liner material

min-max: 1000.0 - 4500.0 m/s / 3280.8 - 14764

ft/s

X6.9

liner thickness

min-max: 0.100 - 20.00 mm / 0.004 - 0.787 inch

X6.10

fluid

water, alkanes, alcohols, oil, acids, CxHx

refined, CxHx light, refrigerant, solvents,

caustic soda, other

X6.11

VoS fluid

min-max: 500 - 2500 m/s / 1640.4 - 8202.1 ft/s

X6.12

density

min-max: 0.1000 - 5.0000 kg/l / 6.2428 lb/ft3 to

312.14 lb/ft3

X6.13

viscosity

min-max: 0.100 cSt to 9999 cSt (mm2/s)

X9.1

transducer set

short code for transducer set,

mentioned on sensor

Ta, Tb, Tc, none

X9.7

actual distance

min-max: -10.00 - +9999 mm / -0.394 - +393.7

inch

X12.1/3/5

Tx serial no.

factory serial number of sensor

Ayy, 5 free units

X12.2/4/6

Tx calibration no.

set calibration number of sensor

according to type sticker

9 free units

A Quick setup

A

A

A

A

quick setup

quick setup

quick setup

quick setup

A1

A1

A1

A1

language

language

language

language

english, german, french, dutch

A2

A2

A2

A2

Tag

Tag

Tag

Tag

unique location in plant

12 free selectable digits

A4

A4

A4

A4

analog outputs

analog outputs

analog outputs

analog outputs

only active if HART device

current output A, B or C, pulse output A, B or D

A4.1

measurement

value for HART current output

(depends on pipe configuration: 1 or 2 pipes)

volume flow, mass flow, VoS, flow speed, gain,

SNR, diagnosis value, volume flow 1 or 2, VoS 1

or 2

.book Page 86 Tuesday, August 18, 2009 4:51 PM