6.2 Menu structure

6.2.1 Quick setup

A1

language

>

english / german /

french

↑ ↓

^

A2

Tag

>

fill in using ↑ ↓ >

^

A3

reset

>

A3.1, A3.2,…

↑ ↓

A3.1

reset errors

yes/no

↑ ↓

^

A3.2

counter 1

yes/no

↑ ↓

^

A3.3

counter 2

yes/no

↑ ↓

(underneath counter becomes active if modular IO)

A3.4

counter 3

yes/no

↑ ↓

^

(end)

A4

analog outputs

>

A4.1, A4.2,…

↑ ↓

A4.1

measurement

>

select from list

using ↑ ↓ >

^

use at all outputs

↑ ↓

yes/no

^

if no:

only HART current

output is selected

if yes:

all analog outputs

are selected

A4.2

unit

>

select from list

using ↑ ↓ >

^

A4.3

range

>

fill in using ↑ ↓ >

^

use at all outputs

↑ ↓

yes/no

^

if no:

only HART current

output is selected

if yes:

all analog outputs

are selected

A4.4

low flow cutoff

>

fill in using ↑ ↓ >

^

use at all outputs

↑ ↓

yes/no

^

if no:

only HART current

output is selected

if yes:

all analog outputs

are selected

A4.5

time constant

>

fill in using ↑ ↓ >

^

use at all outputs

↑ ↓

yes/no

^

if no:

only HART current

output is selected

if yes:

all analog outputs

are selected

A5

digital outputs

>

A5.1, A5.2,…

↑ ↓

A5.1

measurement

>

select from list

using ↑ ↓ >

^

.book Page 74 Tuesday, August 18, 2009 4:51 PM