beautypg.com

Start-up – KROHNE OPTISONIC 6300 EN User Manual

Page 65

background image

START-UP

5

65

OPTISONIC 6300

www.krohne.com

07/2009 - 4000263902 - HB OPTISONIC 6300 R03 en

Mounting the DOWN rail in Z-mode

Mounting the DOWN rail in Z-mode

Mounting the DOWN rail in Z-mode

Mounting the DOWN rail in Z-mode

Set transducer positions for both rails according to the table below.

Measure the outer diameter of the pipe with a measuring band.
For Z-mode, you must install the DOWN rail at the opposite location at the pipe. There are two
possible ways to find the exact location:

1. FIND THE LOCATION WITH A FIXED REFERENCE POINT

Calculate the half of the outer diameter. Mark this 180° alignment line on the pipe.

• Mount the DOWN rail in such a way that the transducer is at the marked location.

2. FIND THE LOCATION WITH THE SUPPLIED POSITIONING TOOL

Advised distance [mm]

Advised distance [mm]

Advised distance [mm]

Advised distance [mm]

Transducer position [mm]

Transducer position [mm]

Transducer position [mm]

Transducer position [mm]

100...250

-65

>250

0

Figure 5-6: Find the opposite location with a reference point

1 Measure the distance between the transducer of the UP rail and the reference point.
2 Add the Advised Distance and mark the location on the alignment line.

.book Page 65 Tuesday, August 18, 2009 4:51 PM